IEC 62353 EPUB

Insulation Resistance Applied Parts to Mains. 8. IEC Leakage Measurements. Method Characteristics. Direct Leakage Provides. IEC Standards for the Safety and. Efficacy of Medical Electrical Equipment . Introduction. It might, perhaps, be a little too self-evident to say that there is no. 7 Sep IEC Medical Electrical Equipment—recurrent test and test after repair of medical electrical equipment—defines the requirements of.

Author: Tojagami Shaktisar
Country: India
Language: English (Spanish)
Genre: Education
Published (Last): 28 December 2016
Pages: 198
PDF File Size: 1.17 Mb
ePub File Size: 7.27 Mb
ISBN: 492-8-88179-625-8
Downloads: 76203
Price: Free* [*Free Regsitration Required]
Uploader: Ararr

Iec 62353 this case the manufacturer should prove and ensure that the allowed limits cannot be exceeded.

Safety testing of medical devices: IEC explained. – PubMed – NCBI

If iec 62353 the flexing, changes in resistance are observed, it iecc be assumed that the protective earth conductor is damaged or the connections are no longer adequate. When using direct current the measurement shall be repeated with opposite polarity.

As far as possible, a consensus is required to harmonise the assessment of the safe operation and testing of me equipment and me iec 62353 whilst respecting local requirements and meeting increasing demands for risk management.

Before iec 62353 was published, some countries had standards for measuring the insulation of me equipment.

An Introduction to IEC 62353

For allowable values, refer to Table 2 or Annex E. IEC specifies three methods— direct, differential, and alternative—that can be iec 62353 to determine the leakage of Uec equipment.

See also lEC All tests shall be performed in such manner that no iec 62353 arise for testing personnel, patients or other individuals. The resistance between the protective earth terminal of the me equipment or me system and protectively earthed accessible conductive parts of the equipment, which may in case of a fault become live shall not exceed mQ.

Insulation resistance measurements shall be performed with V d. The ground bond test proves the integrity of the iec 62353 connection between the ground conductor and any metal conductive iecc, which may become live in fault situations with Class I ifc devices.

If such intervals are not defined by the manufacturer.

With modern electronics and digital technology, the use of higher test currents is regarded as no iec 62353 necessary—a fact recognized by IEC with its mA minimum current. The IEC equipment leakage can be performed using a direct, iev, or alternative method. Care should be iec 62353 to assess the cord and not these connections.

Safety testing of medical devices: IEC 62353 explained.

It is often part of a service carried out by in-hospital mechanical or clinical engineering teams. This edition constitutes a technical revision including iec 62353 following main revisions: Standards are also reviewed periodically; a standard along with amendments is reaffirmed when such review indicates that no changes are needed; if the review indicates that changes are needed, it is taken up for revision.

If iec 62353 of the personnel and of the environment is guaranteed. Enter IEC IEC Medical Electrical Equipment—recurrent test and test after repair of iec 62353 electrical equipment—defines the requirements of ensuring the in-service electrical safety of EM equipment and systems.

Figures 4 and 5 provide a schematic representation of the applied part leakage test on Class I grounded ME equipment.

In addition, when analog instruments were widely used for low-resistance measurement, it was often necessary to use high-test currents iec 62353 produce sufficient voltage drop across the sample to generate iec 62353 necessary needle deflection.

This second edition cancels and replaces the first edition of IEC published in This booklet is written as a guideline for people involved in testing medical electrical equipment ME equipment.

Iec 62353 eic criteria for insulation resistance values used in several countries are mainly based upon iec 62353 from that time. An insulation resistance measurement may be helpful: This residual current is measured by means of an additional winding on a transformer; it corresponds to the leakage iec 62353.

Iec 62353 only exception is made for the equipment leakage current where the allowable iec 62353 would be twice the jA. It also shall be ensured irc they continuously adapt their knowledge to ief current state of the art.

Qualified personnel shall perform these tests. For the purpose of this standard, the application of measuring methods is independent of the edition according to which the ME equipment or ME system is designed.

Therefore the allowable values are twice the values of the other methods.

IEC | IEC Webstore

When determining the correct iec 62353 of testing a specific medical device, ensure that the chosen safety test procedures are applicable to the device under test and are clearly iec 62353 for future use.

In the interests of helping to guarantee safer practice and the iec 62353 of test measurements, IEC defines different types of leakage current tests—one for total equipment leakage and iec 62353 for applied parts leakage currents.

This shall be done by a qualified and authorized individual. Published on September 7, Clause 5 – Tests Clause ie comprises a series of tests, which may be used in testing before 6235 into service, during recurrent testing and in testing after repair. Requirements for supply voltage as defined in lEC do not have to 6233 fulfilled. NOTE To prevent damage to the equipment, a measurement of insulation resistance between applied parts and protective earth connector, respectively enclosure, may only be performed if the equipment is suited to stand such measurement.

This iec 62353 contains iiec with allowable values relating to different editions of IEC iec 62353 Either value measured shall not exceed the allowable value.